3.2 Installation Space

4. Check against Packing List
For composing a liquid chromatograph system with the
ChromasterUltra R
S
and 765 mm deep is required at least. Behind each unit, a space
of 150 mm or more must be secured. At the left and right sides of
the respective units, a space of minimum 100 mm each is
necessary.
For installation, prepare a level and sturdy bench which
withstands a load of 150 kg or more.
The system conceived here is an example in which
ChromasterUltra R
S
Pump, ChromasterUltra R
R
6310 Column Oven, and ChromasterUltra R
S
Array Detector are stacked.
Confirm the necessary space taking the actual system
configuration and OA equipment in the installation room into
account and secure an adequate space at the time of installation.
Check the delivered items against the packing list.