Philips ReCare 7.0 U User Manual page 62

Table of Contents

Advertisement

Immunity test
IEC 60601
test level
Electrical fast
+/-2 kV for power
transient/burst
supply lines
IEC 61000-4-4
+/-1 kV for
input/output lines
Note
• If the device is interfered by power or signal cable, image quality
may be reduced or abnormally displayed. Such kind of interference
images could be easily identi ed and di erentiated from the physi-
ological characteristics of patient and longer clinical time consumed
but wouldn't have any diagnostic accuracy issue.
• If there is a certain frequency of image interference, there is a need
of isolation or ltering of the RF signal.
Compliance level
Electromagnetic environment
+/-2 kV for power
Mains power quality should be that
supply lines
of a typical commercial or hospital
environment.
+/-1 kV for
input/output lines
62