Guidance and manufacturer's declaration –
electromagnetic compatibility
The 7.0 U is intended for use in the electromagnetic environment speci ed below. The customer or the
user of the 7.0 U should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Immunity test
IEC 60601
test level
Electrostatic
6 kV contact
discharge (ESD)
8 kV air
IEC 61000-4-2
Power frequency
(50/60 Hz)
3 A/m
magnetic eld
IEC 61000-4-8
Radiated RF
3 V/m
IEC 61000 -4- 3
80 MHz to 2,5 GHz
Recommended separation distances between
portable and mobile RF communications equipment and the 7.0 U
The 7.0 U is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the 7.0 U can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the 7.0 U as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
150 kHz to 80 MHz
W
0,01
Compliance
Group 1
Class B
Compliance level
6 kV contact
8 kV air
3 A/m
3 V/m
Separation distance according to frequency of transmitter
80 MHz to 800
d = 1,2
MHz
d = 1,2
0,12
0,12
61
Electromagnetic environment –
guidance
The 7.0 U uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The 7.0 U is suitable
for use in all
establishments, including domestic
establishments
Electromagnetic environment
guidance
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic elds
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
d = 1,2
80 MHz to 800 MHz
d = 2,3
800 MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey, a) should
be less than the compliance level in
each frequency range. B) Interference
may occur in the vicinity of equipment
marked with the following symbol:
m
800 MHz to 2,5 GHz
d = 2,3
0,23