9. Specifications
Overall specifications of the LF664 detector and type specification code tables to specify each item
of the detector specifications are described in the following sections.
9.1 Specifications
Meter size: ANSI 150
20", 24" (500, 600mm)
AWWA class D
28", 30", 32", 36", 40", 42", 48", 54", 60", 66", 72", 78"
(700, 800, 900, 1000, 1050, 1200, 1350, 1500, 1600, 1800, 1950mm)
Measurement range in terms of flow velocity: 0–1.0 ft/s to 0–32.8 ft/s (0–0.3 m/s to 0–10 m/s)
Accuracy:
< 20" to 24" (500mm to 600mm) >
± 0.3% of Rate * (Accuracy when combined with the converter LF620F and LF622F)
* This pulse output error result is established under standard operating conditions at Toshiba
admitted flow calibration facility.
* Individual meter's measurement error may vary up to ± 0.5 % of Rate at 3.28 ft/s (1.0 m/s) or
more and ± 0.3% of Rate ±0.079 inch/s (2 mm/s) at 3.28 ft/s (1.0m/s) or less.
* Current output: plus ±8μA(0.05% of span.)
* Refer to individual calibration data for each meter's measurement error.
< 28" and over (700mm and over ) >
± 0.5% of Rate * (Accuracy when combined with the converter LF620F and LF622F)
* This pulse output error result is established under standard operating conditions at Toshiba
admitted flow calibration facility.
* Individual meter's measurement error may vary up to ± 0.8 % of Rate at 3.28 ft/s (1.0 m/s) or
more and ± 0.4% of Rate ±0.157 inch/s (4 mm/s) at 3.28 ft/s (1.0m/s) or less.
* Current output: plus ±8μA(0.05% of span.)
* Refer to individual calibration data for each meter's measurement error.
Fluid conductivity: 5 µS/cm minimum
Fluid temperature: 20" to 40" (500 to 1000mm) ; 14 to 140 °F (–10 to +60 °C)
42" (1050mm) and over ;
Ambient temperature: 14 to 140 °F (–10 to +60 °C)
Fluid pressure: -0.1MPa to 1.0 MPa
Structure: IP 67 and NEMA 4X Watertight
Option 1: IP68 and NEMA 6P Temporary submersible type
The coating for this type is polyurethane coating.
Option 2: IP68 and NEMA 6P Continuous submersible type
The coating for this type is tar epoxy resin coating. This type of flowmeter is sub-mersible
to 5 m in water.
Connection flange standard: ANSI 150, AWWA class D
14 to 104 °F (–10 to +40 °C)
—
—
38