Toshiba LF664 Instruction Manual page 30

Electromagnetic flowmeter detector
Hide thumbs Also See for LF664:
Table of Contents

Advertisement

 
(3) Connecting the input signal cable
Strip the sheath from the end of each conductor of a 2-core individually shielded cable as
shown in Figure 5.4. Twist those shields and cover them with a thermal contraction tube or
vinyl tube not to make contact with the case or core wires. Then attach an M3.5 crimping
terminal with insulated sleeve as shown in Figure 5.3. Connect a crimping terminal to the A
and B terminals on the terminal block and connect to each G terminal of the detector and
converter.
G Shield
B White
                     
     
Figure 5.3 Terminal Treatment of Flow Rate Signal Cable
● Notes on signal cable shield processing work
When stripping an external sheath, intermediate and insulated sheath, be careful not to
scratch or cut the internal conductors and shield mesh. Do not disjoint the shield mesh but
treat it as shown in Figure 5.9.
Shield mesh
       
Figure 5.4 Treating the Signal Cable Shield Mesh
 
A Black
25~45mm
Coated wire
29
6F8A0943
Thermal contraction tube or vinyl tube
a. Open the shield mesh with a pair of tweezers or the like.
b. Pull out the internal coated wires from the hole of the
shielded mesh.
c. Pull out all internal coated wires and extend the shield mesh
wire.