Protocol Set-Up Siemens Balance; Protocol Set-Up Planmed - Siemens NIC2-2 Installation And Startup Manual

Network id camera
Hide thumbs Also See for NIC2-2:
Table of Contents

Advertisement

NIC2-2 · Installation and Set-up Manual
NOTE! The communication protocol has to be set before setting up the picture, otherwise
the COMM field will display NONE.
Example
We want to create a COMM field that shall display the kV. Create a new field and under field-
definition select Type COMM. According to the Siemens Mammomat® 300 protocol the
length of the kV value is five. This means that we have to define the length of this COMM
field to five. Select Comm kV will link the kV data to this field.
NOTE! SIEMENS Mammomat® 300 is sending also the type together with the value, e.g.
'25 kV'..
Fielddefinition
Type: Comm
Row:
Comm: kV

3.4.2.3 Protocol set-up SIEMENS Balance

No set-up is made for this protocol. The data input has the correct format which means that
no translations are needed.
Creating the picture that will be printed on the film is done in the same way as usual selecting
the COMM field type.
NOTE! The communication protocol has to be set before setting up the picture, otherwise
the COMM field will display NONE.
Example
We want to create a COMM field that shall display the kV. Create a new field and under field-
definition select Type COMM. According to the Siemens Balance protocol the length of the
kV value is six. This means that we have to define the length of this COMM field to six. Select
Comm kV will link the kV data to this field.
NOTE! SIEMENS Balance is sending also the type together with the value, e.g. '12.3kV'..
Fielddefinition
Type: Comm
Row:
Comm: kV

3.4.2.4 Protocol set-up PLANMED

No set-up is made for this protocol. The data input has the correct format which means that
no translations are needed.
Creating the picture that will be printed on the film is done in the same way as usual selecting
the COMM field type.