Hitachi EUB-5500 Technical Manual page 45

Ultrasound diagnostic scanner
Table of Contents

Advertisement

   1) RAM Test
Click the "Execute" button to check the memory.
MAP1 and MAP2 PCB. The result of memory check is displayed as "SUCCESS" or
"FAILURE".
After executing memory check, if clicking each button in "Diagnostic Tests" and "Close"
button in "Service Tools", the following message is displayed.
imaging mode, the system needs to be shut down.
the other check programs (Pattern Test and Channel Test) cannot be used.
2) Pattern Test
Press each button in the dialog to perform the following operation. Each button has an image
mode recommendable; therefore, the Test Pattern dialog needs to be displayed after selecting t
Boards to be checked are DBF, EPI,
Also, after executing memory check,
3 - 35
To return to any normal
L1E-EA0229