Step
2
3
4
5
6
7
ÄKTA pilot 600 User Manual 29274559 AA
Action
Disconnect the tubing from the top of the UV monitor and replace it with a
piece of tubing leading to the waste container.
Flush the flow cell with 25 to 30 mL distilled water, using a syringe attached
to the Luer-to-SNAP connector.
Flush the flow cell about five times with 10 mL detergent solution.
Tip:
Warm the detergent solution to 40°C for enhanced cleaning effect.
Allow the detergent solution to remain in the flow cell for about 15 minutes.
Flush the flow cell twice with 25 to 30 mL distilled water.
Disconnect the Luer-to-SNAP connector and waste tubing from the UV
monitor, and reconnect the system tubing.
5.3 Cleaning procedures
5.3.2 Clean the UV flow cell
W
5 Maintenance
99