Drying - Hitachi EUP-B712 Instruction Manual

Biopsy probe
Table of Contents

Advertisement

5.4 Automated cleaning and disinfection
The following items must be provided prior to automated cleaning and
disinfection:
a) Washer disinfector: according to DIN EN ISO 15883 with chemo-thermal
program (temperature: max. 60°C)
b) Detergent: Korsolex® Endo-Cleaner (Bode Chemie; #972020) or another
cleaning agent with approved material compatibility for this medical
device
c) Disinfectant: Korsolex® Endo-Disinfectant (Bode Chemie; #972030)
or another disinfectant with approved material compatibility for
this medical device
d) Washer disinfector accessories
 basket for holding the needle guide bracket
 basket with lid for holding the dismantled small parts (screws,
spring) of the needle guide bracket
1) The parameter of the cleaning and disinfection of the device are
as follows:
Program step water (40 l)
Pre- Rinse
Cleaning
Rinse
Disinfection Deionized water 10 (1 %)
Rinse
Rinse

Drying

2) Place the pre-cleaned and dismantled needle guide bracket and the
dismantled parts (srews, spring) into the basket. Small parts must
be placed in a strainer basket with lid.
3) After closing the door, start the chemo-thermal program.
4) After end of the program, open the door.
5) Remove the dismantled needle guide bracket and check whether it is
dry. If not, proceed as described under drying.
Cold water
Deionized water 5 (0,5 %)
Deionized water
Deionized water
Deionized water
Table 2: Parameters for automated process
Dosage (ml/l) Temp. (°C) time (min)
-35-
5
50
10
1
55
5
1
55
1
55
15
Q1E-EP1412