Hitachi EUP-B712 Instruction Manual page 12

Biopsy probe
Table of Contents

Advertisement

Before use, make sure there is no damage to the surface to which the
needle cannula is attached, and the needle guide works properly.
3) Confirm that the needle guide angle corresponds to the angle
indicated on the monitor. The needle guide angle is engraved on
bracket. (See Fig. 3)
Needle guide
angle
4) Visually inspect the cover to ensure that there are no defects or
holes. Do not use the cover if it has any holes and tears.
2.2.3
Magnetic sensor attachment and Spacer for EZU-RV2S
Visually inspect that the surface of the Magnetic sensor attachment
and
the
Spacer
denaturalization.
Because the Spacer for EZU-RV2S is small, please do not lose the spacer.
CAUTION
CAUTION
21°
Fig. 3 Needle guide angle
for
EZU-RV2S
CAUTION
CAUTION
Needle guide angle
is engraved
for
any
crack,
-6-
deformation
or
Q1E-EP1412