Wall Receptacle Test - GE Dash D 0 0005 G Series Service Manual

Port docking station
Table of Contents

Advertisement

Test Equipment

Wall Receptacle Test

Revision B
Maintenance: Electrical Safety Tests
The manufacturer recommended test equipment required to perform
electrical safety tests is listed below. Equivalent equipment may be
substituted as necessary.
Required Tools/Special Equipment
Item
Leakage Current Tester
120 V (or equivalent)
240 V (or equivalent)
Multimeter
Before starting the tests, the wall receptacle from which the monitoring
device will get electrical power must be checked. This test checks the
condition of the wall receptacle to ensure correct results from leakage
tests.
For international wall receptacles, refer to the internal standards
agencies of that particular country. Use a digital multi meter to ensure
the wall receptacle is wired properly.
If other than normal polarity and ground is indicated, corrective action
must be taken before proceeding. The results of the following tests will be
meaningless unless a properly wired wall receptacle is used.
Dash Port Docking Station
2000966-037
Part Number
DALE 600
DALE 600E