Fig. 11 Connection of cables to the biopsy controller
6.
7.
NOTE
NOTE
8.
9.
Opdima?
Installation
Connect the other end of the cable to the biopsy controller connector X206 in
Fig. 11.
X201
X202
X203
Ferrite
X204
X205
X206
Connect the MAMMOMAT printer (if applicable) to X205 on the biopsy controller
using the previously removed printer cable modified in the following way:
If the shield braid of the printer cable is bare where it was connected to the
generator bottom plate, the cable shall be insulated.
If connected, eject pin 1 on the printer connector side and insulate.
Strain relieve the cable at the bottom of the stand.
The X205 connector is also used for connection of the service PC.
The switch S3 (TEST) on PC Board D702 is not used with Opdima.
The software will sense if a printer or a service PC is connected.
Connect the printer to the same wall socket (branch circuit) as the workstation.
The wall socket must be marked according to Fig. 24.
Close the cable duct cover and install the front cover of the generator.
Register 3
X207
Protective earth cable
SPB7-230.033.10
Rev. 03
08.02
Installation
Mains cable
Page 12 of 26
Siemens-Elema AB
SPS-UD
Solna, Sweden