Protective measures
It is very important that any intervention in the equipment shall start with disconnecting it
from the power supply with the main circuit breaker. To prevent accidental triggering of
high voltage and radiation, set the switch S2 (SS) on board D702 to OFF (lower position,
no triggering of the SS relay).
CAUTION
WARNING
WARNING
CAUTION
Siemens-Elema AB
Solna, Sweden
Shut down the application software before switching off the
power for the workstation (see Logout and turning off the work-
station in Supplement to the Instructions for Use MAMMOMAT
3000 - Opdima?).
Switching off the workstation before the software has been
closed down may cause damage to the files on the hard disc.
If the system is only switched off at the control panel or with S2/
D711 in the MAMMOMAT generator, line voltage will still be
present at the generator line connection, line filter Z1, Z2, trans-
former T1, transformer T10 and board D711 (see MAMMOMAT
1000/3000 Nova Wiring Diagram).
Life-threatening electric shock hazard exists.
Disconnect mains cable and comply with the information on this
page.
After shut-down of the system, there may still be about 380 V DC
present on the intermediate circuit of the MAMMOMAT generator.
Life-threatening electric shock hazard exists.
The voltage level will be indicated by LED V24 on board D710. The
voltage will drop to less than 30 V within about 3 minutes, the LED
goes out at about 30 V.
The boards contain electrostatic highly sensitive components
requiring particular care in their handling.
Risk of damaging components.
Use ESD-equipment, ground prior to making contact and place
the components on a conductive surface, see Technical Informa-
tion TI 219.