Setting Up the Display

Accessing the Display
Submenu
Configuring the Display
Procedure Options
3-2
Complete the following instructions to display the desired data on the
display screen (refer to Figure A-2 on page A-2 in Appendix A, Menu
Maps
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [ ] until PROG appears and press [ENTER].
3. Press [ ] until GLOBL appears and press [ENTER].
4. Press [ ] until I/O appears and press [ENTER].
5. Press [ ] until LCD appears and press [ENTER].
1. At # of LCD PARAMS, press [ ] to select the desired number of
parameters to be sequentially displayed and press [ENTER].
For a 1-channel/path AT868, proceed to Step 3. For a 2-channel/path
meter, proceed to Step 2.
2. Press [ ] to select the desired channel/path option and press
[ENTER].
3. At Measurement Name, press [ ] to select the desired parameter
and press [ENTER].
Note: The measurement units that appear in these prompts are those
selected in the GLOBL-SYSTM menu as described in Chapter
2, Programming Site Data.
The previous two prompts repeat until all of the specified # of LCD
PARAMS have been set up. Proceed to Procedure Options below.
After completing the above steps, the meter returns to the Global I/O
window. Do one of the following:
To perform another programming function, refer to Appendix
A, Menu Maps, to navigate to the desired menu.
To leave the User Program, press [ESC] four times.
After leaving the User Program, the AT868 will reset itself and will
begin to display the parameters specified in this section. If more than
one parameter was set up, each of the parameters will be displayed in
sequence, with a pause of several seconds between display changes.
Displaying Data