Procedure Options
Activating Mass Flow -
MASS
Procedure Options
2-20
After completing the previous steps, the user program returns to the
Advanced Feature window. Do one of the following:
To continue regular programming, refer to Appendix A, Menu
Maps, to navigate to the desired menu.
To leave the user program, press [ESC] five times.
Use this option to calculate mass flow from a static fluid density.
Complete the following steps to enter the static density of the fluid:
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [ ] until PROG appears and press [ENTER].
3. Press [ ] until the desired channel appears. Press [ENTER] at CH1
or CH2.
4. Press [ ] until SETUP appears and press [ENTER].
5. Press [ ] until ADVAN appears and press [ENTER].
6. Press [ ] until MASS appears and press [ENTER].
7. Press [ ] to select whether to activate or deactivate mass flow
(mass flow is calculated from a static density) and press [ENTER].
Note: If you select NO, go to Procedure Options below.
8. Enter the fluid density by using the arrow keys. Press [
to move the cursor to desired location and press
increase or decrease the number and press [ENTER].
After completing the above steps, the user program returns to the
Advanced Feature window. Do one of the following:
To continue regular programming, refer to Appendix A, Menu
Maps, to navigate to the desired menu.
To leave the user program, press [ESC] five times.
]
and [ ]
[ ] and [ ]
to
Programming Site Data