36
3-
3
Inspection and maintenance
3.1

Visual check

SIM
Signs
Check all signs on the system with the model and serial numbers as well as warnings.
The placement of these signs is described in the operating instructions ("Signs" chap-
ter).
Verify that all signs are present and legible.
SI
Customer documentation
The customer must be able to provide the following documents. The documents are to
be checked for availability, completeness, and legibility.
- Operating instructions
- Start-up protocol
- Last maintenance protocol
- Quality assurance protocol
- Maintenance records
- Service reports for repairs and adjustments
- DHHS documentation
3.1.1

Damage

Check the unit for external damage (cracks, breaks, scratches, corrosion etc.).
Check whether the single tank or the collimator show any signs of mechanical defects
that can impair radiation protection.
Repair or replace defective parts.
SIM
Covers
SIM
Detector and detector holder
Check the condition of the detector side covers and replace them if necessary.
SIM
Control panel and display panel
SIM
Release cable
SIM
Single tank
SIM
Collimator
SIM
DAP (optional)
The DAP ionization chamber should not have any scratches.
No damage to DAP supply line, plug or display
SIM
Remote control (optional)
No damage to remote control
MOBILETT XP Digital
SPR8-230.831.30.03.02
11.05
Inspection and maintenance
Page 36 of 74