3
Inspection and maintenance
3.1
Visual check
SIM
Signs
•
Check all signs on the system with the model and serial numbers as well as warnings.
The placement of these signs is described in the operating instructions ("Signs" chap-
ter).
•
Verify that all signs are present and legible.
SI
Customer documentation
•
The customer must be able to provide the following documents. The documents are to
be checked for availability, completeness, and legibility.
- Operating instructions
- Start-up protocol
- Last maintenance protocol
- Quality assurance protocol
- Maintenance records
- Service reports for repairs and adjustments
- DHHS documentation
3.1.1
Damage
•
Check the unit for external damage (cracks, breaks, scratches, corrosion etc.).
•
Check whether the single tank or the collimator show any signs of mechanical defects
that can impair radiation protection.
•
Repair or replace defective parts.
SIM
Covers
SIM
Detector and detector holder
•
Check the condition of the detector side covers and replace them if necessary.
SIM
Control panel and display panel
SIM
Release cable
SIM
Single tank
SIM
Collimator
SIM
DAP (optional)
•
The DAP ionization chamber should not have any scratches.
•
No damage to DAP supply line, plug or display
SIM
Remote control (optional)
•
No damage to remote control
MOBILETT XP Digital
SPR8-230.831.30.03.02
11.05
Inspection and maintenance
Page 36 of 74
CS SD 24
Siemens AG
Medical Solutions