Material Replacement - Siemens MOBILETT XP Digital Maintenance Instruction

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

18
2.3

Material replacement

Material
Collimator lamp
Battery pack (left) BK1/BK2
Battery pack (right) BK3/BK4
Cable winch complete
Option
Remote control batteries
Replace damaged or missing screws.
NOTE
Materials needed for periodic replacement can be requested from the responsible local
Uptime Service Center.
MOBILETT XP Digital
General maintenance information
Only replace damaged or missing screws with steel screws.
All Allen screws must have a tensile strength of 8.8.
SPR8-230.831.30.03.02
11.05
Material no.
08392016
08392024
08392032
06508746
1x9V alkaline
Page 18 of 74
CS SD 24
Interval
12 months
24 months
24 months
48 months
12 months
Siemens AG
Medical Solutions

Hide quick links: