RHYTHMX ECG ANALYSIS PERFORMANCE

The AED RHYTHMx ECG Analysis system analyzes the patient's ECG and advises you when the AED detects a
shockable or non-shockable rhythm.
This system makes it possible for a person, with no training in the interpretation of ECG rhythms, to offer defibrillation
therapy to victims of sudden cardiac arrest.
CARDIAC RHYTHMS USED TO TEST THE RHYTHM RECOGNITION DETECTION SYSTEM
FOR CARDIAC SCIENCE AED
Rhythm Class
Shockable Rhythm – VF
Shockable Rhythm – VT
Non-Shockable Rhythm – NSR
Non-Shockable – Asystole
Non-Shockable– all other rhythms
2023491-201 Rev B
Specifications
Meets AAMI DF 39 requirement and AHA recommendation of Sensitivity of >90%
Meets AAMI DF 39 requirement and AHA recommendation of Sensitivity
of >75%
Meets AAMI DF 39 requirement (>95%) and AHA recommendation (>99%) of
Specificity
Meets AAMI DF 39 requirement and AHA recommendation of Specificity of >95%
Meets AAMI DF 39 requirement and AHA recommendation of Specificity of >95%
© 2007 General Electric Company
SECTION 4 – TECHNICAL DATA