Aed Indicators; Aed Indicators - GE AED Pro Service Manual

Hide thumbs Also See for AED Pro:
Table of Contents

Advertisement

AED INDICATORS

The following indicators are located on the AED.
RESCUEREADY® STATUS INDICATOR
The STATUS INDICATOR is located on the AED handle. When this indicator is GREEN, the device
is RescueReady. This means the Responder AED Pro self-tests have verified the following:
When the STATUS INDICATOR is RED, maintenance is required.
AUDIBLE MAINTENANCE INDICATOR
When the daily, weekly or monthly self-test determines service is required, an audible beep is sounded every 30 seconds
until the lid is opened or the battery power is depleted. Opening and closing the lid may deactivate the beep. If the error is
not corrected by the next automatic self-test, the beep will be reactivated.
DIAGNOSTIC PANEL
A
K
E
H
A.
SMARTGAUGE BATTERY Indicator
This indicator displays the battery capacity. At maximum charge, the battery is GREEN. With use, the
GREEN level will gradually go out from right to left as the battery capacity decreases. Once the battery
level is depleted, the battery indicator will turn to RED and flash, and the battery should be replaced.
Note: When the battery indicator is initially RED – upon lid opening or at any time during a rescue
– a "BATTERY LOW" prompt will be issued at once. However, the AED is capable of delivering at
least nine more defibrillation shocks after the first time a "BATTERY LOW" prompt is issued.
Page 18 of 33
Battery has an adequate charge.
Pads are properly connected to the Responder AED Pro and in working order.
Integrity of the internal circuitry is good.
Note: When Status Indicator is RED or Service Indicator is illuminated, device cannot be
used to perform a rescue.
B C
D
G
G
© 2007 General Electric Company
F, J
2023491-201 Rev B
I

Advertisement

Table of Contents
loading

This manual is also suitable for:

2023681

Table of Contents