portable and mobile RF communications equipment and the GEMS‐H 
The GEMS‐H is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. 
The  customer  or  the  user  of  the  GEMS‐H  can  help  prevent  electromagnetic  interference  by  maintaining  a  minimum  distance  between 
portable  and  mobile  RF  communications  equipment  (transmitters)  and  the  GEMS‐H  as  recommended  below,  according  to  the  maximum 
output power of the communications equipment. 
Rated a maximum 
output power of transmitter [W] 
0.01 
0.1 
10 
100 
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using 
the equation applicable to the frequency of the transmitter. where is the p is the maximum output power rating of the transmitter in watts (W) 
according to the transmitter manufacturer. 
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies 
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and refection from structures, 
objects and people. 
 
Guidance and manufacturer's declaration ‐ electromagnetic immunity 
The  GEMS‐H  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF  disturbances  are  controlled.  Portable  RF 
communications  equipment  should  be  used  no  closer  than  30  cm  (12  inches)  to  any  part  of  the  GEMS‐H.  Otherwise,  degradation  of  the 
performance of this equipment could result. 
Immunity test 
Band a 
Proximity 
fields 
380 ‐ 390 MHz 
from       
RF wireless  
430 – 470 MHz 
Communications 
IEC61000‐4‐3 
704 – 787 MHz 
800 – 960 MHz 
1700 – 1990 MHz 
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11) 
Recommended separation distance between 
Separation distance according to frequency of transmitter [m] 
150 kHz to 80 MHz 
d=1.2 �� 
0.12 
0.38 
1.2 
3.8 
12 
a
Service
 
TETRA 400 
Pulse  modulation 
18 Hz 
GMRS 460 
FM 
FRS 460 
±5 kHz 
deviation  
1 kHz sine 
LTE Band13, 17 
Pulse  modulation 
217 Hz 
GSM800:900 
Pulse modulation  
TETRA 800 
18 Hz 
iDEN 820 
CDMA 850 
LTE Band 5 
GSM 1800 
Pulse  modulation 
CDMA 1900 
217 Hz 
GSM 1900 
DECT 
IEC 60601‐1‐2:2014 
IEC 60601 
Modulation 
test level 
27 V/m 
28 V/m 
9 V/m 
28 V/m 
28 V/m 
80 kHz to 2.7 GHz 
d=2.0 �� 
0.20 
0.63 
2.0 
6.3 
20 
Compliance level 
27 V/m 
28 V/m 
9 V/m 
28 V/m 
28 V/m 