6.8.5. Mitigation of risks due to abnormal actuator module status  
If  the  actuator  module  does  not  work  properly  during  use,  the  controller  may  generate  abnormal  power.  The 
actuator  module  is  monitored  continuously  during  treatment,  and  when  an  abnormality  is  detected,  the  assistive 
force is turned off. 
 
Status of the device and recommended user action:  
Should abnormal actuator status be detected, the tablet PC will display the error "DETECTED ABNORMAL ACTUATOR" 
The actuator status is considered abnormal when data transmitted from the actuator does not match the expected 
signals, for example, when the device detects patient movement, but actuator readings are constant. This may result 
from  disconnected  communication  lines,  actuator  faults,  and/or  software  failure.  The  physical  therapist  should 
attempt to fix the problem by rebooting both the device and the tablet PC. If this corrects the error, the tablet PC will 
display "NORMAL OPERATION". 
If rebooting the device and the tablet PC do not clear the problems, please contact the manufacturer. 
 
Error message examples 
 
 
[Warning message] 
[Screen of normal operation] 
 
 
 
 
 
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11) 
Page 58 of 79