Samsung GEMS-H Instructions For Use Manual page 73

Table of Contents

Advertisement

EMC Table – Guidance and manufacturer's declaration 
The GEMS‐H is intended for use in electromagnetic environments, as specified below.  
The customer or the user of the GEMS‐H should ensure that it is used in such an environment. 
Emission test 
RF emissions 
CISPR11 
RF emissions 
CISPR11 
 
Guidance and manufacturer's declaration ‐ electromagnetic immunity 
The  GEMS‐H  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The  customer  or  the  user  of  the  GEMS‐H  should 
assure that it is used in such an environment. 
Immunity test 
Electrostatic discharge (ESD) 
 
 
 
IEC61000‐4‐4 
Electrical fast  
Transient/burst 
IEC61000‐4‐4 
Surge 
IEC61000‐4‐5 
Voltage dips, 
Short 
Interruptions and voltage  
variations on power supply input lines 
 
IEC61000‐4‐11 
Power frequency (50/60 Hz) magnetic 
field 
 
IEC61000‐4‐8 
NOTE : U
is the a.c. mains voltage prior to application of the test level. 
a For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz. 
 
 
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11) 
Electromagnetic emissions
Compliance
Group 1 
The GEMS‐H uses RF energy only for its internal function. Therefore, its 
RF emissions are very low and are not likely to cause any interference 
in nearby electronic equipment. 
Class A 
The  GEMS‐H  is  suitable  for  use  in  all  establishments  other  than 
domestic,  and  may  be  used  in  domestic  establishments  and  those 
directly  connected  to  the  public  low‐voltage  power  supply  network 
that  supplies  buildings  used  for  domestic  purposes,  provided  the 
following warning is heeded: 
 
Warning! The GEMS‐H is intended for use by physical therapists only. 
This equipment/ system may cause radio interference or may disrupt 
the  operation  of  nearby  equipment.  It  may  be  necessary  to  take 
mitigation  measures,  such  as  re‐orienting  or  relocating  the  GEMS‐H 
or shielding the location.  
IEC 60601 
test level 
±8 kV contact
±8 kV contact
 
 
 
 
 
 
±2,4,8,15 kV air 
±2,4,8,15 kV air 
±2 kV for power 
±2 kV for power 
supply lines 
supply lines 
±0.5 kV, ±1 kV 
±0.5kV, ±1 kV 
differential mode  
differential mode  
 
 
±0.5 kV, ±1 kV, ±2 kV 
±0.5 kV, ±1 kV, ±2 kV 
common mode 
common mode 
0% U
 
0% U
T
(100% dip in U
(100% dip in U
T) 
for 0.5/1 cycles 
for 0.5/1 cycles 
 
 
70% U
 
70% U
T
(30% dip in U
)
(30% dip in U
T
 
for 25/30 cycles 
for 25/30 cycles 
 
 
0% U
 
0% U
T
(100% dip in U
)
(100% dip in U
T
 
for 250/300 cycles  
for 250/300 cycles  
30 A/m 
30 A/m 
 
 
Electromagnetic environment – guidance 
Compliance 
Electromagnetic environment‐ guidance 
level 
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the 
relative  humidity  of  at  least  30%  is 
recommended. 
The  main  power  quality  should  be  that  of  a 
typical commercial or hospital environment. 
The  main  power  quality  should  be  that  of  a 
typical commercial or hospital environment. 
The  main  power  quality  should  be  that  of  a 
T
typical  commercial  or  hospital  environment.  If 
T) 
the  user  of  the  GEMS‐H  requires  continued 
operation  during  power  mains  interruptions.  It 
 
is  recommended  that  the  GEMS‐H  be  powered 
T
)
from  an  uninterruptible  power  supply  or  a 
T
 
battery 
 
T
)
T
 
Power  frequency  magnetic  fields  should  be  at 
levels  characteristic  of  a  typical  commercial  or 
hospital environment 
Page 73 of 79