Electrical Connection - Miele G 7881 Operating Instructions Manual

Dental thermal disinfector
Table of Contents

Advertisement

,
All electrical work must be
performed by a qualified person in
accordance with local and national
safety regulations.
The machine comes equipped for
connection to a 208 V, 60 Hz, 1 phase
power supply. It is equipped with a 6 ft.
long, 10/3 AWG power cord, without a
plug. It must be hardwired into an
electrical junction box that has been
installed on the wall or in a cabinet
adjacent to the machine. It should be
connected to the main power supply.
Voltage:
Frequency:
Rated Load:
Circuit breaker:
Black: connect to L1 (hot)
White: connect to L2 (hot)
Green: connect to GND (ground)
208 V, 1 phase
60 Hz
6 kW
2 x 30 A

Electrical connection

^ Connection should be made via a
suitable isolator, with an "On/Off"
button easily accessible for service
work.
^ A damaged power cord must only be
replaced with a genuine Miele cord
by a Miele service technician.
WARNING
THIS APPLIANCE MUST BE
GROUNDED
Grounding connection
^ The ground lead must be connected
to the screw connection point
(marked with the ground symbol 8) at
the back of the machine.
The machine must only be operated
with the voltage, frequency and
fusing shown on the data plate
located on the rear of the machine,
and on the plinth (behind the service
panel).
The conversion diagram and the
wiring diagram is secured to the inner
side of the service panel.
51

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents