GE LOGIQ E9 Service Manual page 171

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
5-7-7
Shear Wave (SWAVE) Option (cont'd)
SWAVE Acquisition
Pressing the "Start" set key initiates SWAVE acquisition.
Post-SWAVE Acquisition
System displays the acquired SWAVE image and background B-mode image.
User can cycle through the acquired frames, measure and annotate.
SWAVE images can be stored in both raw data and DICOM format.
In order to support the Shear Wave functionality, the Main Power Supply and the GTX Boards have
been modified.
The cover is removed from the Capacitor Pack Module to show the components ONLY. The cover must
NEVER be removed in the field. The Capacitor Pack can store up to 112 Volts.
DANGER
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS
EQUIPMENT. DO NOT REMOVE THE CAPACITOR PACK COVER.
. 2
EV
Figure 5-26 Capacitor Pack Module
Figure 5-27 Capacitor Pack Module LEDs
R3
J2
3
6
J1
4
F1
J3
Chapter 5 Components and Functions (Theory)
BULK CAPACITANCE BOARD
DS1
J4
5487883
A
5487884
REV
x
R1
e
1
BAR CODE
J5
ASSEMBLED IN
LOGIQ E9 S
R4
J6
DS2
R2
e
1
J7
M
ERVICE
ANUAL
5 - 31