GE LOGIC C2 Series Technical Publication page 60

Ultrasound system
Table of Contents

Advertisement

Site Preparations
Electrical requirements for Console Ultrasound Systems (continued)
Table 3-4:
Electrical Specifications for LOGIQ C Series and LOGIQ C3/C5 Premium
Parameter
Voltage Range
Power
Line Frequency
Power Transients
Decaying Oscillation
3-8
The following power line parameters should be monitored for
one week before installation. We recommend that you use an
analyzer Dranetz Model 606-3 or Dranetz Model 626:
Area
100-120V~
220-240V~
All applications
All applications
All applications
All applications
Ultrasound System
500VA
500VA
MAX. 750 VA
50/60Hz (±2Hz)
Less than 25% of nominal peak
voltage for less than 1 millisecond
for any type of transient, including
line frequency, synchronous,
asynchronous, or aperiodic
transients.
Less than 15% of peak voltage
for less than 1 millisecond.
Common Service Information
Direction 5444964-100 English Rev. 5
Limits