GE LOGIC C2 Series Technical Publication page 110

Ultrasound system
Table of Contents

Advertisement

Care and Maintenance
Lead to lead leakage test record (continued)
NOTE:
Table 4-12:
Unit under test____________________________________
Test Conditions
System Power
Off
closed
Off
open
On (Transmit)
closed
On (Transmit)
open
NOTE:
Table 4-13:
Unit under test____________________________________
Test Conditions
System
Power
Grounding/PE
off
closed
off
open
on
closed
on
open
off
closed (reversed
polarity)
off
open (reversed
polarity)
on
closed (reversed
polarity)
on
open (reversed
polarity)
4-32
Not all test procedures are applicable to all areas of the world.
Reversed Polarity testing content satisfies regions following IEC
62353:2007 and IEC 60601-1:2005.
Typical data format for recording patient lead to lead leakage
Grounding/PE
Values in italics font are given as examples only.
Typical data format for recording patient lead to lead leakage
Limit
µA
10
50
10
50
10
50
10
50
Ultrasound System
Patient Lead or Combination Measured
RA to LA
LA to LL
Patient Lead or Combination Measured
RA to LA
LA to LL
Common Service Information
Direction 5444964-100 English Rev. 5
Date of test:_____________
LL to RA
Date of test:_____________
(RA+LA+
LL to RA
LL) to
GRND

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents