GE LOGIC C2 Series Technical Publication page 119

Ultrasound system
Table of Contents

Advertisement

Isolated Probe (Sink) Leakage-Isolation Test (continued)
NOTE:
Table 4-18:
Unit under test___________________________
Test Conditions
System Power
off
closed
on
closed
NOTE:
Table 4-19:
Unit under test____________________________________
Test Conditions
System
Grounding/
Power
PE
off
closed
on
closed
off
closed
(reversed
polarity)
on
closed
(reversed
polarity)
Ultrasound System
Direction 5444964-100 English Rev. 5
Record all data and keep the record of the result with other hard
copies of maintenance data.
Not all test procedures are applicable to all areas of the world.
Reversed Polarity testing content satisfies regions following IEC
62353:2007 and IEC 60601-1:2005.
Typical data format for recording Isolated Probe (sink) leakage
Grounding /PE
Probe names in italics font are given as examples only.
Typical data format for recording Isolated Probe (sink) leakage
Limit
µA
5000
5000
5000
5000
Common Service Information
Date of test:___________
Probe as measured in saline bath
4C
i12L
Probe as measured in saline bath
4C
i12L
Electrical safety tests
TS
E8C
Date of test:_____________
TS
E8C
4-41