Combination operation
3.2 Operating modes
The mounting position of the sample gas restrictor can be derived from the operating modes
selected for the specific application.
3.2.2
Operating modes: Selection of operating mode
Selection of operating mode
The operating modes selected for the specific application result in the following consequences:
● The mounting positions of sample gas restrictors or clamping screws at the sample gas
connections must be checked and changed, if necessary. Additional information is available
at: → Conditions for use/mounting (Page 29).
● The parameter assignment of the gas path must be adapted accordingly. Additional
information is available in the operating manuals: → Table A-3 References 3 - Operating
Instructions (Page 115).
● It may be necessary to adjust the signal frequency of the module combination accordingly.
Additional information is available in the operating manuals: → Table A-3
References 3 - Operating Instructions (Page 115).