Diagnostic Tests
Firmware Versions
Language:
Serial Number:
Options:
Shocks:
2-16
Diagnostic Tests
The M3500B/M5500B includes an extensive set of diagnostic tests, which
test the major hardware components of the defibrillator.
The System Log
These instructions describe how to print the system log, which includes the
unit's serial number, hardware configuration, and a listing of error codes. The
System Log should be printed each time a Performance Verification Test is
run.
1. Enter Diagnostic Mode as described in "Diagnostic Mode" on page 2-11.
2. Select Print Log and press
3. The printer will print the System Log strip.
4. Check the printout to verify the printed results are consistent with the
hardware in place. Check options installed (SpO
serial number (on the bottom of the case). If the printout is not correct,
investigate and resolve the source of the mismatch. Then reset the hard-
ware options and serial number as needed using the Support Software
Tool. See "The Language Support Tool" on page 4-67.
5. If there are errors reported in the System Log:
Check the time and date stamps to see if they are recent errors or not.
a.
Consult Table 3-2 "Error Codes" on page 3-11 to identify the errors.
Then begin troubleshooting as needed (see the "Troubleshooting"
chapter.)
Check that the time and date are correct. If they are not correct, access
b.
Configuration Mode and reset them.(If there are no errors, the time
and date are not displayed.) See "Configuration Mode" on page 2-10.
Figure 2-2 Sample System Log Printout
SYSTEM LOG
Main:
34
DSP:
02
196:
57.00
Key:
06
SpO2:
02.42 01.04
English
US01000241
Pacer
SpO2
2
.
ENTER
Error Codes
90007 18:33
19 Jan 2000
Performance Verification and Safety Tests
, pacing) and the unit's
2