Philips HEARTSTART XLT Service Manual page 198

Advertisement

Specifications
Table 7-1 Delivered Energy vs. Load Impedance
Selected
Energy
(J)
25
5
4.7
10
9.3
25
23.4
50
46.7
70
65.4
100
93.5
150
140.3
200
187
7-2
Delivered Energy (J)
Load Impedance (ohms)
50
100
5
5.2
10
10.4
25
26.2
50
52.3
70
73.1
100
104.7
150
156.8
200
209.3
Charge Time: Less than 3 seconds to 200 Joules with a new, fully charged
M3516A SLA battery pack at 25
powered by a M3517A or M3518A power module with no battery installed.
Patient Impedance Range: 25 to 180 Ohm.
Manual Mode
Manual Output Energy (Delivered): 5, 10, 25, 50, 70, 100, 150, 200 Joules.
Controls: On/Energy Select, Off, Charge/Disarm, Shock, ECG Lead Select,
SpO
, HR Alarms, Exit Manual Mode, Sync, Pacer Start/Stop, Pacer Rate,
2
Pacer Current, Pacer Mode, Display Contrast, ECG Gain, Volume, Print Strip,
Print Summary, Annotated Mark Event.
Indicators: LCD display for ECG waveform and text prompts, Audio alerts,
QRS Beeper, Charging Tone, Charge Done Tone, Manual Mode LED, Sync
LED, Pacer LED, Printer.
Armed Indicators: Charge done tone and available energy indicated on dis-
play.
Energy Selection: Front Panel "1" key.
Charge Control: Front Panel "2" key.
Shock Control: Front Panel "3" key.
Synchronizer: SYNC message appears on the monitor and is annotated on
the printer (if printing while in Sync Mode). An audible beep sounds with
each detected R-wave, while a tick mark on the monitor and printed strip indi-
cate the discharge points. Synchronizer delay is less than 60 msec from peak
R-wave to peak current of the defibrillation discharge.
125
150
5.4
5.2
10.7
10.4
26.9
26.2
53.5
52.1
75.0
72.9
107.2
104.4
161.0
156.5
214.6
208.6
o
C. Less than 15 seconds to 200 Joules when
Accuracy
+ 2 J
+ 2 J
+ 4 J
+15%
+15%
+15%
+15%
+15%
Specifications