Device Command List For The Pulse Signal - HP 81110A Quick Start Manual

Pulse/pattern generators
Table of Contents

Advertisement

Getting Started
Examples

Device Command List for the Pulse Signal

# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlay OFF
# Pulse stream operating mode is required, but as we start from a
# default status it is not necessary to send a command for setting the
# instrument into pulse stream operating mode.
# Set the period to 20 ns, the pulse width to 10 ns, the leading edge to
# 3 ns and the trailing edge to 5 ns.
:PULSe:PERiod 20NS
:PULSe:WIDTh1 10NS
:PULSe:TRANsition1 3NS
:PULSe:TRANsition1:TRAiling:AUTO OFF
:PULSe:TRANsition1:TRAiling 5NS
84
If you want to include this Pulse Signal in your program use the fol-
lowing command lines in boldface font. The comment lines starting
with a # are not required.
The commands are listed with the channel specifier for channel 1, if
there is a second channel installed and you want to program it with a
similar signal, use the same commands and replace 1 by 2 for the
second channel, where required.