Installation and Calibration
Note
Note
32
7. If any status is \Failed" highlight ADJUST 5
COMPLETE AUTOMATIC ADJUSTMENTS (1, 2,
3, 4) and press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT ITEM
set up a shown in gure 1-8..
The automated adjustments may take up to two hours
per band to complete. If your Fader is option 147, it
may take four hours to complete automated adjustments.
8. When the automated adjustments are nished,
highlight TEST 04 COMPLETE PERFORMANCE
TESTS (01, 02, 03) and press
9. Check the PASS/FAIL STATUS printed at the
bottom of each printout page and the summary
sheet.
10. If the tests now pass, you may want to store your
new calibration data in EEPROM. Highlight STORE
ADJUSTMENT DATA IN NON-VOLATILE
EEPROM MEMORY and press
11. If the Performance Test printout still indicates a
failure, refer to chapter 3 of the Service Manual,
\Troubleshooting".
Figure 1-11 shows the sequence of tests necessary after
repair of the Multipath Fading Simulator.