HP 37718A Installation And Verification Manual page 158

Hide thumbs Also See for 37718A:
Table of Contents

Advertisement

Performance Tests
SDH Transmit/Receive Jitter Amplitude Accuracy
3 Select TRANSMITTER
OUTPUT SETTINGS
and setup the display as shown
opposite.
4 Select RECEIVER
INPUT SETTINGS
and setup the
JITTER
display as shown
opposite.
5 Set the Calibration Pattern to ALT.
6 Set the Spectrum Analyzer as follows:
Centre Frequency - 77.76 MHz (half bit rate being tested)
Frequency Span - 30 kHz (approx 3 times the Jitter frequency)
Reference Level - approx -15dBm
Sweep Time - Auto
Video Bandwidth - Auto
Resolution Bandwidth - Auto
7 Set the Synthesizer to Frequency 10 kHz and minimum Output level.
8 Increase the Synthesizer output level until the Carrier and first sidebands are
equal as observed on the Spectrum Analyzer (see Figure 3-11).
3-118
JITTER

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omniber 719Omniber 71837719a

Table of Contents