Recommended Test Equipment - HP 5065A Operating And Service Manual

Rubidium vapor frequency standard
Hide thumbs Also See for 5065A:
Table of Contents

Advertisement

Model 5065A
Circuit Diagrams, Theory, and Maintenance
Table 5-5. Recommended Test Equipment
Instrument
Primary
Frequency
Standard
Frequency
Stability Analyzer
system
50-ohm Feedthru
Termination
RMS Voltmeter
Oscilloscope
Spectrum
Analyzer
Vector
Voltmeter**
Strip Chart
Recorder
RF Voltmeter
DC Power Supply
Time Interval
Counter
DC Electronic
Voltmeter
Phase
Comparator**
Clip-on DC
Milliammeter
Variable
Line Source
Wave Analyzer
Required Characteristics
Frequency: 5 MHz and 1 MHz
Output Level: 1V rms at 50 ohms
Accuracy: ±1 x 1 0 - n
Capable of automatically measuring
short term frequency fluctuations with
averaging times of 10 msec and greater
50 ohms shunt connections male and
female BNC
Voltage Range: .3V to 3V full scale
Frequency Range: 10 Hz to 10 MHz
Accuracy: ±5% full scale
Vertical Frequency Response: dc to 50 MHz
Sensitivity: .005V/cm
Calibrated Sweeps: 2 sec to .05/usec/cm
Frequency Response: 1 kHz to 110 MHz
Response: ±0.5 dB
Sensitivity: -130 dBm
Scan Width: 2 kHz to 100 MHz
Stability: Residual FM less than 20 Hz
peak-to-peak
Frequency: 1 MHz to 1 GHz
Voltage Range: 1.5 mV to 1V rms
Chart Speed: 1, 2, 4, 8, in./hr.
Spans: . 1 , .5, 1, and 5V full scale
Input Resistance: 200k/volt
Accuracy: .2% full scale
Range: 10 mV to 10V rms
Frequency Range: 500 kHz to 60 MHz
Accuracy: ±3 full scale
Range: 0 to 20 Vdc
Output I: 0 to 1A
Line Regulation: 0.001%
Resolution of >2 nsec
per measurement
Resistance Range: 10H to 10 M f l
Voltage Range: 0.1 to 100V full scale
Voltage: ±2% full scale
Frequency Range: 100 kHz to 10 MHz
Input Sensitivity: .1Vrms
Output: 1V into 100Kfl for 360° phase change
Range: 3 mA to 30 mA
Accuracy: ±0.1 mA ±3% full scale
Variable from 103V rms to 127V rms and
206V rms to 254V rms
B.W. 10 Hz
Frequency: 137 Hz
Sensitivity: 100 nV
Use
Performance Check
Performance Check
Performance Check
Troubleshooting
Performance Check
Troubleshooting
Performance Check
Troubleshooting
Adjustments
Performance Check
Performance Check
Performance Check
Performance Check
Troubleshooting
Troubleshooting
Performance Check
Performance Check
Troubleshooting
Adjustments
Performance Check
Troubleshooting
Troubleshooting
Troubleshooting
Model
HP 5061A
Option 004
HP 5390A
Opt 010 with
HP 1051A
(Option H66)
HP11048B
HP 3400A
HP 180 with
H P 1 8 2 0 C a n d
HP1805A
HP 10006A Probe
HP 8552B and
HP 8553B
HP 8405A**
HP 680A
HP411A
HP6101A
HP 5370A*
HP410C
K34-59991A**
HP 428B
Superior electric
powerstat
(115V line)
3PF116or
(230V line) 3PF216
HP3581A
*Not needed if HP 5390A Frequency Stability Analyzer available.
"The Model 8405A or the Model K34-59991A may be used for frequency offset and comparison measurements.
It is not necessary to have both equipments.
5-11