Alarms 4001 - 4124 EZChrom

GCP Alarm Descriptions 4001 - 4124 EZChrom
The following tables list the alarm number (#), type (+ information, ? warning, ! error) alarm
text, description, and actions.
#
Text
4001
? TFTP cannot load with ac‐
tive method %3
4003
!
Zero Correction Out of Lim‐
its on %4
4022
? I/O channel not found on
%4
4024
? Slope check failure on
channel %4
4025
? Detector channel under‐
flow occurred on %4
4026
? Detector channel open oc‐
curred on %4
4027
? Detector channel overflow
occurred on %4
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
4.6 Alarm Codes, Descriptions, and Suggested Actions
Description
SNECON PICs: The content of the status
buffer, 'Status_data', was invalid and not
recognized by the PIC. This may happen
if the PIC firmware is out of date relative
to the NS486 software.
SNECON PICs: The value of the sta‐
tus_type was invalid and not recognized
by the PIC.  This may happen if the PIC
firmware is out of date relative to the
NS486 software.
SNECON PICs: A packet was received
with an invalid I2C message size or no
I2C message at all. The packet was dis‐
carded.
SNECON PICs: The PIC lost a slave
read arbitration. It indicates that at least
one other module responded to the same
message request.
Description
Method currently running when software
load was commanded.
System error.
Hardware resource requested not
present.
Slope check commanded with invalid pa‐
rameters.
A/D converter for detector reading lowest
possible value. This error may also be a
secondary error caused by an overflow
on an FID DPM.
Open connection detected on A/D detec‐
tor input. This error may also be a secon‐
dary error caused by an overflow on an
FID DPM.
A/D converter for detector reading maxi‐
mum value possible.
General Maintenance and Troubleshooting
Action
Reload the OS and APP software for
the SNECON.
Replace the SNECON.
Reload the OS and APP software for
the SNECON.
Replace the SNECON.
Reload the OS and APP software to all
SNECONs.
If sporadic, no action is necessary.
If repeating frequently, reset the ana‐
lyzer.
Action
Put method on hold and wait for it to
complete before loading SNE software.
Contact Customer Support.
Reload and or repair SYSCON data‐
base.
Check method for slope check events
and review data.
Check A/D inputs, potentially replace
referenced DPM.
Verify proper operation of the DPM.
Check A/D inputs, potentially replace
referenced DPM.