General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
#
Text
427
+ Detector simulation activa‐
ted
428
!
Invalid configuration for
smoothing noise measure‐
ment
460
!
Invalid Method Write
461
!
Argument %3 , Invalid
Type %4
462
!
Invalid Method Section %3 System error
463
!
Unable to Run Method,
Hardware Initializing
464
!
Unable to Run Method, In‐
strument Busy
465
!
Error Running Method
466
!
Error Installing Method
467
!
Error Retrieving Method
468
!
Unable to Run Method, Not
on Method List
469
!
Invalid component results
470
!
Invalid Spectrum results
471
!
Invalid calibration file %3
line %4
472
? Unknown method status
%3
473
? Multiple component sets
not allowed
474
? Multiple component scans
not allowed
475
!
Component report invalid
476
!
Arguments do not match
script
477
!
Component report with no
associated method
478
!
Internal reset commanded
40
Description
The detector data reported by the DPM
is from a simulation chromatogram read
from the EZChrom trace binary and not
from actual data acquired by the DPM.
This is normal when running a simulation.
The method contains a channel with
smoothing noise measurement and the
time range used for noise measurement
is outside of the time range of the chro‐
matogram.  The analysis was aborted
and the results for this channel could not
be calculated.
System error
System error
System error
System error
An I/O error was detected while execut‐
ing a cycle event associated with a run‐
ning method. This is probably caused by
a missing or invalid I/O module or by re‐
using the same detector channel in more
than one application.
System error
System error
System error
System error
System error
System error
System error
System error
System error
System error
System error
System error
System error
Action
If simulation is not desired, the trace
(TRC) binary must be removed from the
EZChrom method for non-simulated
chromatogram data to be collected.
Correct the method (on the workstation)
so that the time range used for noise
measurement falls within the time
range of the chromatogram.
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Verify the status of all I/O modules and
correct as needed. To identify the I/O
module associated with this alarm, read
the alarm cycle time information and
use while looking up the list of cycle
events in the method. The problematic
I/O is one of the events with the cycle
time matching the cycle time reported
with the alarm.
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001