Alarms 330 Through 359 Sne Communication - Siemens Maxum II Manual

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

4.6.2

Alarms 330 through 359 SNE Communication

GCP Alarm Descriptions 330 - 359
The following tables list the alarm number (#), type (+ information, ? warning, ! error) alarm
text, description, and actions.
#
Text
331
!
Run Method: No SNE
found or bad status on mod‐
ule: %3
332
!
Run Method: No module
found for detr: %3
333
!
Run Method: No Detr found
or bad status for meth‐
od.channel: %3
334
!
Run Method: No Channel
found for method: %3
335
+ SNE connection opened
on %3
336
!
SNE connection closed on
%3 error: %4
337
!
SNE connection replaced
on %3
338
? SNE %3 Method %4 Can‐
not Store Chrom
339
? SNE %3 Method %4 Appli‐
cation %5 not found
340
!
SNE %3 Method %4 Load
- Invalid Method
341
!
SNE %3 Method %4 Inac‐
tive
342
!
SNE %3 Method %4 Load
- Max Method exceeded
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
4.6 Alarm Codes, Descriptions, and Suggested Actions
Description
The connection between the GC and the
SNE is invalid.
The SNE has not reported the detector,
pressure controller, or temperature con‐
troller.
Realtime chromatogram attempt on inva‐
lid detector, or bad status on detector.
No channels are present or can't find de‐
tector for channel.
System error
SNE connection closed due to timeout or
error.
System error
Results received from SNE for unknown
stream.
Can't find application or method to match
SNE results. This indicates that messag‐
es between the SNE and SYSCON are
corrupted.
SNE has sent a status message that the
method is invalid. No other information is
available.
SNE sent message that method is inac‐
tive. No other information is available.
SNE sent message that maximum meth‐
ods has been exceeded.
General Maintenance and Troubleshooting
Action
Check cable between the SNE and SY‐
SCON.
Check the LEDs on the SNE to see if it
is running.
Check hardware connections to SNE.
The GC can contain references to inva‐
lid hardware channels.
Check the pressure, temperature, and
detector channels defined in the appli‐
cation for correct assignments.
Check hrdwr_id, module for app_detec‐
tor. The GC can contain references to
invalid hardware channels.
Check the detector channels defined in
the application for correct assignments.
Check app_detector, EZChrom method
for proper hardware channel assign‐
ments.
Contact Customer Support.
If IP address specified is not a
192.168.144.# network address, check
for appropriate grounding of system.
Otherwise check SNE for appropriate
connections and software versions.
Contact Customer Support.
Check stream table. It is possible to de‐
lete streams during the run of a cycle.
If that is done, then this alarm may oc‐
cur.
Reset SNE to sychronize messages. It
is possible to delete applications during
the run of a cycle. If that is attempted,
then this alarm may occur.
Download method from EZChrom
again.
Restart the application.
Reduce number of methods, reset SY‐
SCON.
33