Print Report window*
Use this window to print sample data using report layouts created in the Print Format
Settings window.
This window is only available from the supervisor logon. To print reports, go to the
Review/Edit window and using the Print Report button.
CAUTION
Tests on the Print Report form must be in the same order as they appear in the System
Test List window, to avoid incorrect positioning of results on the print report. Do not
modify the order of the test names on the System Test List window after the initial
setup. Any repositioning of test names on the System Test List window could result
in already run being associated with the wrong test name.
Test Select window*
You select tests in this window for patient and control sample analysis and calibrations.
Use this window for the following tasks:
•
Disable ("down") a test.
For example, a test may be downloaded from a host system to analyzers in your lab,
but that test is not run on the ADVIA Chemistry System.
•
To "down" a test, select one of the Test Select radio buttons, and deselect it from the
corresponding test table.
You can remove tests for calibrators and controls from the current run using the
NOTE:
Calibration
Cup/Tube Assign window*
Use this window to assign container types to each of the 84 positions in the STT sample
tray. For the current analysis, you can override the settings in the Temp cup/tube select
window.
Test Utilization Statistics window
Use this window to monitor the test utilization against Patient, Calibrations, and Controls.
All tests defined on the system are displayed.
NOTE:
44
and
windows, but they stay selected for subsequent runs.
Control
ADVIA 2400 Operator's Guide