Cup/Tube Assignment - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

Similarly, you can download workorders at any time as a batch using the Order Entry
window.
3. Use the Item Setting dialog box at the Online Setting window to determine which test
results to transmit and to assign host test numbers.
4. Use the Data Clean Setting dialog box at the Online Settings window to have the
system automatically validate the sample results before they are transmitted.
Use the Transfer Results List dialog box of the Review/Edit to review the data clean
report for each sample.
5. In the System basic composition area at the System Specification Settings window,
select
6. In the During-operation system set area of the System Monitor window, select
the Online option.
7. Use the Host button in the Operation Panel to disconnect from or reconnect to the
host computer.

Cup/Tube Assignment

Assign container types to each of the 84 positions in the STT sample tray. For the current
analysis, you can override the settings in the Temp cup/tube select window.
1. Log on as
2. At the Menu Panel, select
3. At the Cup/Tube Assign window, assign the containers as follows:
a. Select the arrow in the Status/Container field for a position.
b. From the list, select a container type to assign it to the position.
CAUTION
c. As an option, you can select I for priority analysis of the sample at this position.
d. Repeat for other STT positions as needed.
4. To save the changes you made, select
5. When prompted, select
6. To save your changes and close the window, select
7. When prompted, select
System Setup
.
Avail
.
supervisor
Request
Container Types are defined in the System Specifications Settings window.
If you want to specify the container type in the workorder, keep the
Status/Container set to 0:Priority requested (the other Status/Container values
have priority over the workorder settings).
When sliding the cover, be careful not to scratch the tubes and dilution bowl.
Also, when fastening the screw, verify that the cover is not caught in the groove
and is not loose.
Yes
.
Yes
, then select
Cup/Tube Assign.
.
Save
.
.
Exit
for
Do
175