Recommended Separation Distances - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Recommended separation distances

Rated maximum output
power of transmitter in watts
0.01
0.1
1
10
100
a
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2001989-351A
The following table provides the recommended separation distances (in meters)
between portable and mobile RF communications equipment and the ApexPro
Telemetry Transmitter.
The ApexPro and CARESCAPE Telemetry Transmitter is intended for use in the
electromagnetic environment on which radiated RF disturbances are controlled. The
customer or the user of the ApexPro Telemetry Transmitter can help prevent
electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the ApexPro and
CARESCAPE Telemetry Transmitter as recommended below, according to the
maximum output power of the communications equipment.
Separation distance in meters (m) according to frequency of transmitter
a
150 kHz to 80 MHz
d = 11.6
1.16
3.67
11.6
36.7
116
For transmitters rated at a maximum output power not listed above, the recommended
separation distance [d] in meters (m) can be estimated using the equitation applicable
to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE
These guidelines may not apply in all instances. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
ApexPro™
a
80 MHz to 800 MHz
d = 11.6
1.16
3.67
11.6
36.7
116
Electromagnetic compatibility
a
800 MHz to 2.5 GHz
d = 23.3
2.33
7.37
23.3
73.7
233
B-5