Performance Specifications; Physical Specifications; Certification; Antenna Specifications - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Transport and storage conditions

Physical specifications

Subsystem

Certification

Receiver Unit
Subsystem

Antenna specifications

Performance specifications

2001989-351A
Temperature
Relative humidity
Height
Width
Depth
Weight
420-474 MHz – R&TTE
560 – 614 MHz – FCC Part 15, Subpart B Class B (U.S. only)
UL 60601-1 Classified. IEC 60601-1 and IEC 60601-1-2 Certified. CE marked per
the Medical Devices Directive 93/42/EEC. Yakuji (Japanese Ministry of Health,
Labour and Welfare).
Type
(-002, -004)
Voltage range
7.0 – 12.6 V
Minimum voltage
7.0 V
Current draw
55 mA
Gain
17 dB
ApexPro™
–40°C to 70°C (–40°F to 158°F)
15% to 95% (non condensing)
170 mm (6.7 in)
325 mm (12.8 in)
250 mm (9.8 in)
6.4 kg (14 lb) with 4 quad receiver modules
Hi-Pwr
Passive
(-003, -005, -006)
N/A
N/A
N/A
–5dB
Technical specifications
Active
Active
(-007)
(-008)
8.0 - 12.6 V
8.0 - 12.6 V
8.0 V
8.0 V
115 mA
52 mA
28.5 dB
15 dB
A-13