Physical Specifications; Fcc Compliance Information; Certifications - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Physical specifications

FCC compliance information

Certifications

2001989-351A
Graph request
Event Marker
Maximum transmitters
Dynamic range
Input offset
Input impedance
ECG gain selection
ECG gain accuracy
Common mode rejection
Defibrillator protection
Defibrillator recovery time
Pacemaker detection
Patient leakage current
Serial communications
Height
Width
Depth
Weight
This device complies with Part 95 of the FCC Rules.
Operation of this equipment requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry Service.
UL/IEC/EN 60601-1
IEC/EN 60601-1-1
IEC/EN 60601-1-2
IEC/EN 60601-1-4
IEC/EN 60601-2-27
IEC/EN 60601-2-49
1395.025 to 1399.975 MHz - FCC Part 95
ApexPro™
Transmitted
Transmitted
199 active within WTMS at a single facility
± 5 mV (RTI)
± 300 mV (RTI)
15 M ohm minimum differential @ 10 Hz
5, 10, 20, 40 mm/mV (RTI)
± 5% @ 15 Hz
100 dB minimum @ 60 Hz
± 5000 VDC, 360 joules into 100 ohm
Transmitter recovers within 5 seconds
± 2 mV to ± 700 mV (RTI); 100 µsec to 2 msec; either
polarity; on multiple leads
Meets UL/IEC 60601-1
2 ports at 9600 baud asynchronous
13.7 cm (5.38 in)
7.3 cm (2.875 in)
2.3 cm (0.91 in)
141.8 g (0.275 lb) without battery; 170.1 g (0.375 lb) with
battery