Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)
3
Record the actual reading in the Performance Test Record
(Test 5, Line 1).
4
Change the synthesized signal generator's frequency to 1.5 GHz.
Verify the Counter reads 1.5 GHz ±1 Hz
5
Record the actual reading in the Performance Test (Test 5, Line 2).
6
Disconnect the test setup.

Minimum to Maximum Accuracy Test for Options

030/050/124
Equipment
HP 8663A Synthesized Signal Generator (not needed for Option 050)
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector (not needed for Option 050)
HP 10503A BNC Cable (2)
HP 10100C 50Ω Feedthrough
HP 11500D N-N Cable (Option 050 only)
HP 1250-1250 N(M) to SMA(F)
HP 1250-0777 N(F) to N(F)
Counter Setup
1
Cycle the POWER key to preset the Counter.
2
Set the Counter as follows:
MEASURE/Freq & Ratio, FREQUENCY 2
Gate & ExtArm, TIME: TIME 1.000 s