To 3.0 Ghz Sensitivity For Option 030 Only - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
3
Increase the power level in 5 dB steps until the Counter displays a
stable reading of approximately 2.5 GHz.
The Counter should display 2.5 GHz at ≤ −27.0 dBm (10 mVrms).
4
Sweep the frequencies from 2.5 GHz to 200 MHz.
The Counter should read frequencies from 2.5 GHz to 200 MHz at
≤ −27.0 dBm (10 mVrms).
5
Record the actual reading in the Performance Test Record
(Test 5, Line 1).

2.7 to 3.0 GHz Sensitivity for Option 030 only

1
Disconnect the HP 8663A Synthesized Signal Generator from the
Counter (see Figure 1-9), and connect the HP 8340B Sweep
Oscillator to Channel 3 of the Counter.
Change the settings of the Sweep Oscillator to 3.0 GHz, −40 dBm.
2
3
Increase the power level in 1 dB steps until the Counter displays a
stable reading of approximately 3.0 GHz.
The Counter should display 3.0 GHz at ≤ −21.0 dBm.
4
Sweep the frequencies from 3.0 GHz to 2.7 GHz.
The Counter should read frequencies from 3.0 to 2.7 GHz at ≤ −21.0 dBm.
5
Record the actual reading in the Performance Test Record
(Test 5, Line 2).
6
Disconnect the test setup.
1-32
Assembly-Level Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53181a53131a

Table of Contents