HP 35601A Operating And Service Manual page 21

Spectrum analyzer interface
Table of Contents

Advertisement

Model 35601A
Installation
PIN
1
2
3
4
13
14
15
16
5
17
6
7
8
9
10
11
12
18
19
20
21
22
23
24
L
M#£h&fl-E*ft#frS
J
LINE
DI01
0102
0103
DI04
0105
DI06
DI07
0108
EOI
REN
DAV
NRFD
NDAC
IFC
SRQ
ATN
SHIELD-CHASSIS GROUND
P/O TWISTED PAIR WITH PIN 6
P/O TWISTED PAIR WITH PIN 7
P/O TWISTED PAIR WITH PIN 8
P/O TWISTED PAIR WITH PIN 9
P/O TWISTED PAIR WITH PIN 10
P/O TWISTED PAIR WITH PIN 11
ISOLATED DIGITAL GROUND
The -hp- 35601A
contains metric threaded HP-IB cable
mounting studs as opposed to English threads. Metric
threaded -hp- 10631 A, B, C or D HP-IB cable lockscrews
must be used to secure the cable to the instrument. Iden-
tification of the two types of mounting studs and
lockscrews is made by their color. English threaded
fasteners are colored silver and metric threaded fasteners
are colored black. DO NOT mate silver and black fasteners
to each other or the threads of either or both will be
destroyed. Metric threaded HP-IB cable hardware illustra-
tions and part numbers follow.
LOCKSCREW
LONG MOUNTING STUD
SHORT MOUNTING STUO
1390-0360
0380-0643
0380-0644
THESE PINS
ARE
INTERNALLY
GROUNDED
o:
6 . 5 mm
_ i
ff
===:
T)
JL
V
Figure 2-6. HP-IB Connector
2 11- OPERATING AND STORAGE ENVIRONMENT
To prevent potential electrical shock or fire hazard, do not expose
equipment to rain or moisture.
2-12. Operating Environment
In order for the -hp- 35601A to function properly, the operating environment must be within
the following limits:
Temperature
0
°C to +55°C (+32°F to + 131°F)
Relative Humidity
<95%
Altitude.
< 4 6 0 0 m e t r e s ( 1 5 0 0 ( ) f
Magnetic Field Strength
<0.05 gauss
2-7