gas lines
(Do not disconnect here.
Disconnect at unions on the
ends of supply lines )
Thermal gasket
(2021263-001)
Setscrew for FID
bottom body
Setscrew
for FID cover
Figure 4-10
1. Remove the four screws that secure the purge tube to the electrodes of the detector.
2. Pull the purge tube about 1 inch away from the detector assembly and carefully remove
the wires from the electrodes.
3. The purge tube with its bushing and gasket, EC cover plate, and igniter assembly will not
be reused. Set these items aside. However, the screws that were used to connect the tube
to the detector will be reused.
4. Remove the setscrew for the FID bottom body to separate the detector heater and mounting
plate.
5. Remove the old heater assembly. (2021266) This will not be reused.
Maxum Edition II Detectors
Service Manual, May 2018, A5E42019847001
FID
cover
Insulation
Insulation
FID
body
Pipe gasket (A5E41666941001)
FID Installation Exploded Diagram
Detector Bushing
Ground strap set screw goes here
Hole for purge filter
Install signal cable so
end is aligned with
igniter cable ends
EC cover plate
(A5E35374578001)
Heater purge tube sealing grommet (Heyco)
Detector bushing (A5E41666802001)
and o-ring (A6X19905099)
Purge filter (A5E41381855001)
and washer (A5E41849023)
Heater purge tube
Heater assembly
FID mounting plate
Heater block mounting screws (4)
Flame Ionization Detector
4.4 Replacing the FID Igniter
Mounting/ground strap
FID purge tube grommet
(A5E35380226)
Igniter
43