Flame Photometric Detector

3.8 FPD Specifications

3. Reinstall the screw to secure the temperature setpoint module stack through the access
hole on the right side of the unit.
4. Plug the mezzanine module onto its connector on the right side of the unit, and reinstall the
two screws ear the top edge.
5. Insert the rear lug into the slot at the rear of the DPM mounting position with the mounting
stud at the front inserted through the slot in the DPM bracket.
6. Slide backward to lock the rear mounting lug into the slot.
7. Reinstall the nut on the threaded stud to secure the DPM.
8. Reconnect the I
of the board.
9. Reconnect the RTD temperature input cables (top rear of unit, if present).
10.Reconnect the fIlament detector cable if present.
11.Reconnect the detector signal cable from the mezzanine module.
12.Follow the steps in the
3.8
FPD Specifications
Table 3-2
Flame Photometric Detector (FPD)
Parameter
Target Components
Measurement Type
Smallest Measuring Range (at 60°C oven temperature)
Typical Measuring Range (at 60°C oven temperature)
Dynamic Range
Linear Range
Selectivity
Temperature Range
Ignition type
Electrical Data
30
C/power cable and temperature control output cables from the rear edge
2
General Analyzer Startup Procedure .
Value
Sulfur
Mass
0 to 1 ppm ±5%
0 to 10 ppm ±3% FS
up to
0 to 500 ppm ±2% FS
10
2
70 (70:1) linearized
10
to 1
4
Sulfur to Carbon
105 to 150°C
Glow Plug
2V at 3A (Maximum, only for flame ignition)
Maxum Edition II Detectors
Service Manual, May 2018, A5E42019847001