Flame Photometric Detector
3.5 Base3 Detector Personality Module (DPM)
The method is the part of the application that contains the parameters for controlling the
hardware used by one cycle clock. It provides peak areas and component concentrations and
includes all cycle clock timed events. There is one cycle clock per method.
Figure 3-5
Part Number
The Base3 DPM (Part Number A5E02645925001) is shipped with current analyzers. It can
be used as a replacement part for earlier DPMs in Maxum I analyzers using an adapter, part
number A5E34938458001.
24
Status
LEDs
Base3DPM With Mezzanine Module
Access hole for mounting
screw for Temp Setpoint modules
Mezzanine
Module
Mounting
Location
RTD Temp Control Input
Flame Sense and
Filament Detector Input
Ignite Signal/
Igniter Power
Signal Cable
from Detector
Maxum Edition II Detectors
Service Manual, May 2018, A5E42019847001
Location
ID Switch