"Read Instruction Manual" Do Not operate this oxygen concentrator without first reading and understanding this manual. If you are unable to understand the warnings and instructions, contact your local SeQual dealer or authorized service representative before attempting to use this oxygen concentrator.
This instruction manual will inform you about the use and care of the Regalia Oxygen Concentrator and will serve as a reference as you use your oxygen concentrator.
"Alternating Current" Oxygen Concentrator Information SeQual's Regalia Oxygen Concentrators are quiet, reliable and low maintenance sources of oxygen. The Regalia is designed and constructed to consistently provide high concentrations of oxygen in a lightweight, compact cabinet that is easy to maneuver.
DISS fitting and application. Shorter tubing lengths will be required in some cases. The internal diameter of the tubing should be no less than 3/16-inch (0.48 cm). SeQual Technologies Inc. Part No: 5167...
Check that the oxygen concentrator is not placed near a heat register. Important: SeQual’s products operate safely when used according to their marked electrical ratings and product usage instructions. The power cord used with the Regalia Oxygen Concentrator must meet the requirements of the country where the product is used.
5. Flow Rate Adjustment Adjust the flow rate to desired setting. Press the Plus button to increase or the Minus button to "+" " " decrease the flow rate until the proper level is displayed. SeQual Technologies Inc. Part No: 5167...
Resume use of the Regalia when power is restored. If a power outage occurs, the Regalia will activate both an audible and a visual red light alarm to inform you that the Regalia is not operating. The Regalia should be located where the alarms can be seen and heard.
During System Malfunction condition the Green and Yellow lights may be On or Off. 3. Non-Usage - When not using the oxygen concentrator, press the On/Off button to turn it off. SeQual Technologies Inc. Part No: 5167...
Do Not operate the oxygen concentrator for more than 30 minutes without a filter installed. Oxygen Delivery Accessories - Follow the instructions supplied by the manufacturer. SeQual Technologies Inc. Part No: 5167...
Service required. Flow Rate not visible on LCD malfunction Service required. Flow Indicator Display Audible Alarm Remove power cord from wall Red Light On System malfunction outlet. Flow Indicator Blank Service required. Back Light On SeQual Technologies Inc. Part No: 5167...
**Where equipped. Subject to ±5% system concentration accuracy, and for backpressures up to 7kPa. Outlet pressure may decline to 5.0 psig (34.5 kPa) nominal at 10,000 ft altitude. The system performance may decline above 10,000 ft (3048m). SeQual Technologies Inc. Part No: 5167...
Disposal of Equipment and Accessories Follow local governing ordinances and recycling plans regarding disposal of device components. Authorized European Representative: Manufacturer: MeddiQuest Limited SeQual Technologies Inc. Business & Technology Centre 11436 Sorrento Valley Road Bessemer Drive San Diego, CA 92121 USA...