Model 1000 User Manual Contents Important Safety Instructions Symbols Used in User Manual and on Oxygen System Indications For Use Contraindications Safety Guidelines Oxygen Therapy Safety Operating Safety Power Cartridge Safety Introduction To Your ECLIPSE Oxygen System Understanding the Controls and Front Panel Description of the ECLIPSE Control Panel Features User Controls...
Under the Following Headings: SeQual ® Technologies Inc. (SeQual) recommends that this User Manual be given to each patient receiving an ECLIPSE Oxygen System. This User Manual describes all the necessary information in order for a patient or caregiver to operate the ECLIPSE.
Page 5
Model 1000 User Manual The use of oxygen therapy “No Smoking or Naked Flames”: requires that special care be taken to reduce the risk of fire. Materials that burn in air, and even some materials that will not burn in air, are easily ignited and burn rapidly in high concentrations of oxygen.
Page 6
Symbols (continued) “Alternating Current” An audible indicator is provided to announce a “Audible Alarm”: reduction in oxygen concentration or when the power has been interrupted or lost. Turns the unit ON and OFF but does not directly “ON/OFF”: disconnect the mains power. Increases the flow setting in 0.5 “Increase Flow Setting”: increments each time the button is pushed.
Model 1000 User Manual Indications for Use THE ECLIPSE OXYGEN SYSTEM IS INDICATED FOR THE ADMINISTRATION OF SUPPLEMENTAL OXYGEN. THE DEVICE IS NOT INTENDED FOR LIFE SUPPORT NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES. YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS.
“Read Instruction Manual”: without first reading and understanding this manual. If you are unable to understand the warnings and instructions, contact your authorized SeQual dealer/provider before attempting to use this oxygen system. In certain circumstances, oxygen therapy can “Caution or Attention”: be hazardous;...
Your oxygen system should be located in a well-ventilated area to allow for adequate air intake and to avoid the intake of airborne pollutants and fumes. ONLY use this device with accessories designed or recommended by SeQual Technologies Inc. If your oxygen system has been dropped, damaged or exposed to water, contact your dealer/provider for examination or repair.
Page 10
Protect electrical cords from sharp edges. Electrical shock and serious WARNING physical injury may result. Use only the SeQual AC Power Supply and DC Power Supply supplied WARNING with your ECLIPSE Oxygen System. Use of a non-SeQual Power Supply may be hazardous, cause damage to your oxygen system and will void the warranty.
Model 1000 User Manual Power Cartridge Safety This Power Cartridge may explode and cause potential injury WARNING if exposed to or disposed of in a fire. short-circuit the Power Cartridge’s metal contacts with WARNING DO NOT metallic objects, such as keys or coins. It may cause sparks or excessive heat.
Introduction to Your ECLIPSE Oxygen System This User Manual will inform you about the use and care of the ECLIPSE Oxygen System and its accessories. Thoroughly read all the information in this manual in its entirety before operating your ECLIPSE. Contact your dealer/provider for proper training in the use of this device.
Model 1000 User Manual Understanding the Controls and Front Panel Description of the ECLIPSE The key features of the ECLIPSE Oxygen System and control panel are important items to become familiar with. Handle Control Panel Oxygen Outlet Port Service Port (Not for Patient Use) Simplified User Instructions...
Understanding the Controls and Front Panel (continued) Control Panel Features The ECLIPSE displays important operating information using the icons, numbers and visual indicators on the control panel. This section will help you understand these features. No Smoking Icon ON/OFF Button ON/OFF Indicator (Green) Alert Indicator (Yellow ) Flow Rate Setting (2 Digits )
Model 1000 User Manual User Controls This section describes the control buttons and function for your ECLIPSE Oxygen System. F R O N T PA N E L I C O N S This button turns the system ON or ON/OFF Button and Indicator: OFF.
Page 16
Understanding the Controls and Front Panel (continued) Troubleshooting section for the proper response. A flashing yellow indicates a medium priority alert. Prompt response is needed to deal with a condition. Refer to the Troubleshooting section. ALARM (Red) Indicator - High Priority Alarms: Indicates a high priority alarm condition.
Model 1000 User Manual Recommended Operating Environments The following information recommends certain boundaries that will allow the ECLIPSE Oxygen System to operate at its full potential. Operating Temperature 50º F to 104º F (10º C to 40º C) Operating Humidity 10% - 95% at an 82.4°F (28°C) dew point Transport/Storage Temperature -4º...
Recommended Operating Environments (continued) The Proper Location Select a location that avoids fumes and pollutants and allows the oxygen system to draw in room air through the air inlet filter on the back of the cabinet and exhaust air out the vent at the bottom of the case without being restricted. Locate the unit such that the alarms can be heard.
Operating Instructions Before Operating Your SeQual dealer/provider or authorized service representative will demonstrate how to operate the oxygen system. This manual serves as a supplemental reference to help you operate and maintain the unit. If you have any further questions or problems call your SeQual dealer/provider or authorized service representative.
Operating Instructions (continued) Powering Your ECLIPSE System The ECLIPSE is a lightweight, portable oxygen system capable of being operated directly from three different power sources: ( 1) an AC Power Supply, (2) a DC Power Supply, (3) a rechargeable Power Cartridge. (3) RECHARGEABLE POWER CARTRIDGE (1) AC POWER SUPPLY...
Page 21
Use with grounded, 3-pin receptacle. DO NOT remove ground pin from CAUTION cord. Use only with SeQual supplied cords. DO NOT connect the ECLIPSE to an extension cord or electrical CAUTION outlet controlled by a switch.
Page 22
Operating Instructions (continued) Secure the ECLIPSE and the DC Power Supply in your vehicle and WARNING make sure the air inlet and exhaust airflow are not blocked. This may impact performance. leave the ECLIPSE or DC Power Supply plugged into the WARNING DO NOT vehicle without the engine running or attempt to start the vehicle...
Page 23
Model 1000 User Manual When operating in the 10 AMP switch position, flow rate settings above 2.0 LPM continuous flow will automatically switch to 2.0 LPM continuous flow when the DC Power Supply is connected. The ECLIPSE system can also be (3) RECHARGEABLE POWER CARTRIDGE: powered by a rechargeable Power Cartridge.
Page 24
There are no serviceable parts inside the Power Cartridge. CAUTION DO NOT open. Replace the Power Cartridge with only SeQual approved Power CAUTION Cartridge. The Power Cartridge used in this device may present a risk of fire CAUTION or chemical burn if mistreated.
Page 25
Model 1000 User Manual Typical Power Cartridge Recharge Time The typical time to recharge your Power Cartridge to achieve 80% capacity from a fully discharged Power Cartridge is 1.4 hours to 5.0 hours, dependent upon the flow setting. If the Power Cartridge is too warm, charging will not begin until it sufficiently cools. The Power Cartridge may not charge or the charge time may increase when operating from the DC Power Supply set in the 10 AMP position.
Page 26
DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle CAUTION or in the trunk during a hot day. The ECLIPSE system can only work with a SeQual Power Cartridge. CAUTION Use of another Power Cartridge or Battery may damage the unit, present a risk of fire or explosion and will void the warranty.
Model 1000 User Manual Operating Your ECLIPSE STEP 1: LOCATING YOUR ECLIPSE Place the ECLIPSE in a well-ventilated place so that the air inlet and exhaust have clear access. Position the ECLIPSE so that any alarm can be heard. Check to see if the air inlet filter is in place before using. If the air inlet filter needs to be cleaned, wash with soapy water and allow drying before using.
Page 28
Operating Instructions (continued) STEP 2: TURN ON THE UNIT AND ALLOW TO WARM UP Press and hold the Button for one (1) second to turn “ON/OFF” your ECLIPSE. A brief self-test will happen. Verify that all the indicators are lit and a 3-second buzzer sounds, indicating the system is in working order.
Page 29
Model 1000 User Manual STEP 3: CONNECT THE TUBING TO THE OXYGEN OUTLET AND THE CANNULA PER THE INSTRUCTIONS FROM YOUR DEALER/PROVIDER. Replace the cannula and tubing on a regular basis. Consult your dealer/provider for cannula and tubing cleaning and replacement information. STEP 4: SELECT FLOW MODE Press the button to select the flow delivery mode.
Page 30
Pulse Flow Mode Operation • Pulse Flow Mode should only be used under the direction of a qualified clinician or physician. • When operating in Pulse Flow Mode, an alarm will beep after 15 seconds if inspiration is not detected. If another 15 seconds elapses and no inspiration is detected, the unit will switch to Continuous Flow Mode at the last Continuous Flow Mode rate setting used.
Page 31
Model 1000 User Manual STEP 6: BEGIN USE Breathe normally through the cannula. You will get longer operating time on your Power Cartridge if you operate your ECLIPSE in Pulse Flow. STEP 7: CART OPERATION The Cart is designed for use on pedestrian surfaces. Align the locator holes on the bottom of the unit to the tabs on the cart.
Alarm Conditions and Troubleshooting The table shows possible alarm conditions and suggests appropriate responses. If any other alarms occur, call your authorized SeQual dealer/provider. Green Yellow Audible Condition Indicator Indicator Indicator Alarm What to do “Warming Up. Please Wait. ”...
Page 33
Model 1000 User Manual Symptom Probable Cause Remedy ECLIPSE does not power on No Power Cartridge installed Install charged Power Cartridge when ON/OFF button is pressed or plug into external power Power Cartridge is discharged Plug into external power Malfunction Contact your dealer/provider No Oxygen Blocked or kinked tubing...
If your physician has prescribed a Humidifier Adapter - SeQual Part Number 3509: humidifier, you may need to use the SeQual Humidifier Adapter. Follow the instructions for use. Attach the Humidifier Adapter to the oxygen outlet port of the oxygen system and then to the humidifier.
Page 35
Model 1000 User Manual ® Oxygen Tubing - Salter Labs Series 2000 or equivalent tubing (internal diameter Connect the oxygen tubing to of the tubing should be no less than 3/16” (0.48 cm): the outlet port of the humidifier or directly to the oxygen outlet port if you do not use a humidifier.
Page 36
Accessories (continued) Your AC Power Supply enables the ECLIPSE to be AC Power Supply, Model 6104: connected to a grounded AC electrical outlet with the country-specific power cord. Please contact your dealer/provider for the country-specific power cord. Your DC Power Supply enables the ECLIPSE to be DC Power Supply, Model 6105: operated in an automobile or other vehicle with an accessory outlet.
Model 1000 User Manual Cleaning, Care and Routine Maintenance Routine Maintenance A trained technician will carry out all routine maintenance at least once a year or as needed. Your dealer/provider will perform inspections and maintenance of the compressor, filters, alarms, 9V battery, Power Cartridge and other internal parts as needed. disassemble the ECLIPSE or attempt to perform any routine WARNING DO NOT...
Page 38
The filter should be free of liquid water and dry to the touch before CAUTION reinstallation, as excess moisture may impair the proper operation of the oxygen system. When the filter is removed, the Service Port is exposed. DO NOT CAUTION attach any cables to this Service Port.
Page 39
Power Cartridge Cleaning, Care and Maintenance The Power Cartridge in the ECLIPSE Oxygen System requires special care to assure a longer life and higher performance. The SeQual Power Cartridge is the only approved Power Cartridge to be used. Once a month, completely discharge your Power Cartridge and fully recharge to calibrate the Power Cartridge.
Technical Data Flow Rate Settings 1 to 6 setting in Pulse Flow Mode 0.5 to 3.0 LPM in Continuous Flow Mode Oxygen Concentration 90% +/– 3% for all flow settings Oxygen Concentration Indicator Green Light = Normal Operation Yellow Light = Warning or Caution, less than 85% +/– 3% Red Light Flashing = Abnormal Operation, less than 70% +/–...
Page 41
Model 1000 User Manual Pulse Flow Mode Principle of Operation Pulse Flow Mode is designed to deliver 16 ml boluses of oxygen per incremental flow setting for each patient inspiration detected. Bolus volume is held constant regardless of breath rate within a given setting.
Summary of All Warnings and Cautions FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE WARNING ORDER OF A PHYSICIAN. THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE WARNING SUSTAINING APPLICATIONS, NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES.
Page 43
Protect electrical cords from sharp edges. Electrical shock and WARNING serious physical injury may result. Use only the SeQual AC Power Supply and DC Power Supply supplied WARNING with your ECLIPSE Oxygen System. Use of a non-SeQual Power Supply may be hazardous, cause damage to your oxygen system and will void the warranty.
Page 44
Use of a damaged Power Cartridge may cause personal injury. WARNING Exposing the Power Cartridge to water or other liquids may cause WARNING personal injury. Protect the ECLIPSE and AC/DC Power Supplies from fluid spills or WARNING drips to avoid possible shock hazards. DO NOT use any cord if damaged.
Page 45
Model 1000 User Manual Operating your ECLIPSE outside the recommended operating WARNING environment may impact performance, cause damage and will void the warranty. Pulse Flow settings should be determined for each patient individually. WARNING Settings from Continuous Flow applications may not be applicable to Pulse Flow Mode.
Page 46
If tubing becomes disconnected during operation, ECLIPSE does WARNING not alarm. Check flow at the cannula to verify oxygen delivery. DO NOT tamper with, disassemble, crush or heat the Power Cartridge WARNING above 140° F (60° C). The Power Cartridge may present a risk of fire or explosion and will void the warranty.
Page 47
Model 1000 User Manual Availability of an alternate source of oxygen is recommended in case CAUTION of power outage or mechanical failure. DO NOT expose unit to temperature outside of specified operating or CAUTION storage temperatures as this may damage the unit. Locate oxygen tubing and Power Supply cords to prevent tripping CAUTION hazards.
Page 48
There are no serviceable parts inside the Power Cartridge. CAUTION DO NOT open. Replace the Power Cartridge with only SeQual approved Power CAUTION Cartridge. The Power Cartridge used in this device may present a risk of fire or CAUTION chemical burn if mistreated.
Page 49
DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle CAUTION or in a trunk during a hot day. The ECLIPSE system can only work with a SeQual Power Cartridge. CAUTION Use of another Power Cartridge or Battery may damage the unit, present a risk of fire or explosion and will void the warranty.
Consult the manufacturer or service technician for help. Guidance and Manufacturer’s declaration - electromagnetic emissions The SeQual Technologies ECLIPSE Oxygen System is intended for use in the electromagnetic environment specified below. The customer or the user of the SeQual Technologies ECLIPSE Oxygen System should assure that it is used in such an environment.
Fax: + 44 1763 222049 DEALER/PROVIDER NAME AND ADDRESS: PHYSICIAN NAME AND ADDRESS: SeQual ® and ECLIPSE are trademarks of SeQual Technologies Inc., San Diego, CA 92121 ® Salter Labs is a registered mark of Salter Labs, Arvin, CA 92303 Lysol ®...
Page 52
ECLIPSE Model 1000 User Manual PN 6112 03/06...